When a medical device study qualifies for an abbreviated Investigational Device Exemption (IDE), the sponsor generally does not need to submit an IDE application to the U.S. Food and Drug Administration (FDA). The sponsor must explain why the study poses nonsignificant risk (NSR), obtain approval from the reviewing Institutional Review Board (IRB), and meet the other abbreviated IDE requirements in 21 CFR 812.2(b).
Use the guidance and tools below for each role you hold in the study. You may serve in more than one role.
Sponsor
The sponsor initiates the study and takes responsibility for its overall planning and oversight. The sponsor may be an individual investigator or an organization, such as a device manufacturer. Before the study begins, document who will serve as the sponsor so the team knows who is responsible for meeting the abbreviated IDE requirements.
Study examples
The following examples illustrate uses a study team might evaluate. The risk determination depends on the device’s proposed use in the specific study.
- Studying a new use for an FDA-cleared device, such as evaluating a hemodynamic monitoring system for home use when its cleared use is in a clinical setting.
- Evaluating the safety or effectiveness of a product for a medical purpose, such as a recreational mobile app being studied for a medical use, a smartwatch pulse oximetry feature, or a light box being studied as a treatment for depression.
Sponsor resources
- FDA guidance on significant risk and nonsignificant risk medical device studies
- Abbreviated IDE Sponsor and Investigator Checklist
- Abbreviated IDE Monitoring Checklist
- Abbreviated IDE Checklist Presentation (PowerPoint)
Investigator
The investigator conducts the clinical investigation and directs the use of the study device with participants. When a team conducts the investigation, the investigator is the person responsible for leading that team.
Investigator resources
- Abbreviated IDE Sponsor and Investigator Checklist
- FDA guidance on significant risk and nonsignificant risk medical device studies
Device Manufacturer
If a University of Pittsburgh or UPMC laboratory manufactures a device that will be evaluated for safety or effectiveness, review the design and development requirements that apply to the device.
Devices subject to design and development requirements
Under 21 CFR 820.10(c), design and development requirements apply to manufacturers of Class II and Class III devices and the following Class I devices:
- Devices automated with computer software
- Tracheobronchial suction catheters (21 CFR 868.6810)
- Non-powdered surgeon’s gloves (21 CFR 878.4460)
- Protective restraints (21 CFR 880.6760)
- Manual radionuclide applicator systems (21 CFR 892.5650)
- Radionuclide teletherapy sources (21 CFR 892.5740)
Documenting and maintaining design procedures
University-based device manufacturers should establish and maintain procedures to ensure the device meets its design requirements. As part of that work:
- Document design and development processes and their related procedures.
- Review and approve procedures before implementing them.
- Put approved procedures into practice and follow them.
- Review procedures periodically to confirm they still apply.
- Update procedures as needed and review and approve revisions.